Model number/catalog number: 72400024,
serial number: n/a,
batch/lot number: 1100009968,
model/catalog description: balloon.Model number/catalog number: 72404127,
serial number: n/a,
batch/lot number: 1100013046,
model/catalog description: pump.The device is not available for analysis; therefore, no physical or visual analysis of the product could be performed.There was no report of a device performance allegation.The device history record (dhr) review confirmed that the device met all material, assembly and performance specifications.A review of the device instructions for use (ifu) was completed and did not reveal any evidence of device misuse, off label use, or failure to follow instructions.A review of the ams 800 hazard analysis was completed and confirmed that the events of urinary incontinence and atrophy were defined in the risk documentation.This event type has been accounted for during product risk analysis to support acceptable risk benefit for the product.Based on the information available, a conclusion code of adverse event related to patient condition was assigned to this investigation.This record was identified during a retrospective review of complaint events; an electronic medical device report (emdr) was not previously submitted per applicable reporting criteria.Boston scientific has opened a nonconforming events and prevention investigation (b)(4) to determine the potential cause for why an emdr was not previously submitted to the food and drug administration (fda) in accordance with boston scientific's established medical device reporting process and fda adverse event reporting requirements.Information gained through these investigation efforts will be utilized to take action(s), as necessary.
|