• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: BOSTON SCIENTIFIC CORPORATION AMS 800 Artificial Urinary Sphincter; Device incontinence mechanical/hydraulic

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BOSTON SCIENTIFIC CORPORATION AMS 800 Artificial Urinary Sphincter; Device incontinence mechanical/hydraulic Back to Search Results
Model Number 72404134
Medical Device Problem Code Adverse Event Without Identified Device or Use Problem (2993)
Health Effect - Clinical Codes Muscle Weakness/Atrophy (1967); Urinary Incontinence (4572)
Date of Event 12/01/2025
Type of Reportable Event Serious Injury
Event or Problem Description
It was reported that the patient with this artificial urinary sphincter (aus) experienced recurring urinary incontinence and cuff atrophy.As a result, the device was explanted and replaced.No further patient complications were reported.
 
Additional Manufacturer Narrative
Model number/catalog number: 72400024, serial number: n/a, batch/lot number: 1100009968, model/catalog description: balloon.Model number/catalog number: 72404127, serial number: n/a, batch/lot number: 1100013046, model/catalog description: pump.The device is not available for analysis; therefore, no physical or visual analysis of the product could be performed.There was no report of a device performance allegation.The device history record (dhr) review confirmed that the device met all material, assembly and performance specifications.A review of the device instructions for use (ifu) was completed and did not reveal any evidence of device misuse, off label use, or failure to follow instructions.A review of the ams 800 hazard analysis was completed and confirmed that the events of urinary incontinence and atrophy were defined in the risk documentation.This event type has been accounted for during product risk analysis to support acceptable risk benefit for the product.Based on the information available, a conclusion code of adverse event related to patient condition was assigned to this investigation.This record was identified during a retrospective review of complaint events; an electronic medical device report (emdr) was not previously submitted per applicable reporting criteria.Boston scientific has opened a nonconforming events and prevention investigation (b)(4) to determine the potential cause for why an emdr was not previously submitted to the food and drug administration (fda) in accordance with boston scientific's established medical device reporting process and fda adverse event reporting requirements.Information gained through these investigation efforts will be utilized to take action(s), as necessary.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
AMS 800 Artificial Urinary Sphincter
Common Device Name
Device incontinence mechanical/hydraulic
Manufacturer (Section D)
BOSTON SCIENTIFIC CORPORATION
300 boston scientific way
marlborough MA 01752
Manufacturer (Section G)
Boston Scientific Corporation
300 boston scientific way
marlborough MA 01752
Manufacturer Contact
farshad fahimi
4100 hamline avenue north
building c
saint paul, MN 55112
MDR Report Key50694853
Report Number2124215-2026-48199
Device Sequence Number9684669
Product Code EZY
UDI-Device Identifier00878953003122
UDI-Public(01)00878953003122(17)221013(10)1000471880
Combination Product (Y/N)Y
Initial Reporter StateMI
Initial Reporter CountryUS
PMA/510(K) Number
P000053
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2022
Device Explanted Year2026
Reporter Type Manufacturer
Report Source Health Professional,Company Representative
Initial Reporter Occupation Physician
Type of Report Initial
Report Date (Section B) 09/09/2026
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Lay User/Patient
Device Expiration Date10/13/2022
Device Model Number72404134
Device Catalogue Number72404134
Device Lot Number1000471880
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 03/05/2026
Initial Report FDA Received Date09/09/2026
Date Device Manufactured10/13/2020
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Hospitalization; Required Intervention;
Patient Age79 YR
Patient SexMale
-
-