The customer reported that filters would "explel out with coughing & sometimes while just breathing" and have had to change very frequently, nearly every hour to 2 hours; per the customer there was "no choking or falling into airway.No medical emergency." medline quality received and inspected two photos which appeared in used condition and the left side of the filter when looking at the photo was no longer in place.As a result, one side of the filter has come out was confirmed, however, medline was unable to determine a root cause without a physical sample.A retrospective query on the item number for the past six months found no similar complaint(s), the manufacturing site was notified of the issue, and the division will monitor the complaint during quarterly trending.There were no reports of death, serious injury, medical intervention, or follow up care.It has been determined that the reported event could cause or contribute to serious injury if it were to occur again.In an abundance of caution, this is a reportable event.If additional information becomes available this report will be reopened and reevaluated.This medical device report (mdr) is being submitted as part of retrospective review activities, which include review of historical data in accordance with regulations at 21 cfr part 803.The information included in this mdr reflects the information reasonably known to the manufacturer at the time of this retrospective review and submission.
|