The reported product problem/issue was leaked drain bag.Medline quality received a sample for investigation; there was no description of the issue sent with the sample, and the product was received in unpackaged, used condition.Visual assessment of the provided product revealed no abnormalities or defects, measurement of the catheter met all specified dimensional criteria, and functional testing showed the catheter drained all 300ml with no issue; after drainage, a leak was found in the drain bag of the urine meter component, originating from a small hole near the bottom of the drain bag.The reported problem was confirmed, and the cause was changed to dc: unknown responsible party - cannot determine cause w/received sample; based on the findings of the quality investigation, no change was made to the regulatory reportability decision for the complaint.There were no reports of death, serious injury, medical intervention, or follow up care.It has been determined that the reported event could cause or contribute to serious injury if it were to occur again.In an abundance of caution, this is a reportable event.If additional information becomes available this report will be reopened and reevaluated.This medical device report (mdr) is being submitted as part of retrospective review activities, which include review of historical data in accordance with regulations at 21 cfr part 803.The information included in this mdr reflects the information reasonably known to the manufacturer at the time of this retrospective review and submission.
|