Catalog Number 09230815190
Medical Device Problem Code Non Reproducible Results (4029)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 07/15/2026
Type of Reportable Event Malfunction
Event or Problem Description
The initial reporter received questionable online tdm methotrexate assay results for several patient samples tested on the cobas c 503 analytical unit and the cobas pure c 303 analyzer.The following are examples of discrepant results for 3 patients from the list provided: patient 1 sample a on (b)(6) 2026: the initial result was 1.21 umol/l with a data flag.The repeat result was 60.5 umol/l with a data flag.Sample b on (b)(6) 2026: the initial result was 1.21 umol/l with a data flag.The repeat result at reduced sample volume was 16.8 umol/l.Sample c on (b)(6) 2026: results from the c 503 analyzer: the initial result was 1.21 umol/l.The repeat result in reduced sample volume was 16.8 umol/l.This patient sample was rerun on a cobas pure analyzer: the repeat results of a manual 3-fold dilution and an automated 5-fold dilution were approximately 96.3 µmol/l.Repeat results of 32.1 umol/l and 19.1 umol/l were also provided.On (b)(6) 2026: the repeat result at a dilution ratio of 50 was >60 umol/l.The repeat result at a manual dilution ratio of 50 was 92 umol/l.Sample d the result at a manual dilution ratio of 50 was 63.3 umol/l.Patient 2 sample a on (b)(6) 2026: the initial result was 1.21 umol/l with a data flag.The repeat result at reduced sample volume was 16.8 umol/l.The repeat result was 58.6 umol/l with a data flag.Sample b on (b)(6) 2026: the initial result was 1.20 umol/l.The first repeat result was 1.20 umol/l with a data flag.The second repeat result in reduced sample volume was 16.7 umol/l.On (b)(6) 2026: the third repeat result was 1.2 umol/l the fourth repeat result at a dilution ratio of 50 was 58.9 umol/l with a data flag.-- the fifth repeat result at a dilution ratio of 150 was 75.9 umol/l.Sample c the result at a dilution ratio of 50 was >60 umol/l.Patient 3 sample a on (b)(6) 2026: the initial result was 1.21 umol/l with a data flag.The repeat result at reduced sample volume was 16.8 umol/l.Sample b on (b)(6) 2026: the initial result was 1.18 umol/l.The first repeat result in reduced sample volume was 16 umol/l.The second repeat result in reduced sample volume was 16 umol/l.The third repeat result at a dilution ratio of 50 was 15.7 umol/l with a data flag.
 
Additional Manufacturer Narrative
The cobas c 503 analytical unit serial number is (b)(6).The cobas pure c 303 analyzer serial number and the online tdm methotrexate reagent lot number used for this analyzer were not provided.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ONLINE TDM Methotrexate
Common Device Name
Enzyme immunoassay, methotrexate
Manufacturer (Section D)
Roche Diagnostics
9115 hague road
indianapolis IN 46250 -045
Manufacturer (Section G)
Roche Diagnostics GmbH
sandhofer strasse 116
mannheim (baden-wurttemberg) 68305
GM   68305
Manufacturer Contact
natalie catton
9115 hague road
indianapolis, IN 46250
MDR Report Key50695561
Report Number1823260-2026-03678
Device Sequence Number18563238
Product Code LAO
UDI-Device Identifier07613336224823
UDI-Public07613336224823
Combination Product (Y/N)N
Initial Reporter CountryJA
PMA/510(K) Number
K233454
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source Foreign,Health Professional,User Facility
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial
Report Date (Section B) 09/09/2026
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Catalogue Number09230815190
Device Lot Number90676301
Was Device Available for Evaluation? Yes
Is the Reporter a Health Professional? Yes
Initial Date Received by Manufacturer 08/13/2026
Initial Report FDA Received Date09/09/2026
Was Device Evaluated by Manufacturer? (Y/N) Yes
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Patient SexUnknown
-
-