| Catalog Number |
09230815190 |
| Medical Device Problem Code |
Non Reproducible Results (4029)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
07/15/2026
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
The initial reporter received questionable online tdm methotrexate assay results for several patient samples tested on the cobas c 503 analytical unit and the cobas pure c 303 analyzer.The following are examples of discrepant results for 3 patients from the list provided:
patient 1
sample a
on (b)(6) 2026:
the initial result was 1.21 umol/l with a data flag.The repeat result was 60.5 umol/l with a data flag.Sample b
on (b)(6) 2026:
the initial result was 1.21 umol/l with a data flag.The repeat result at reduced sample volume was 16.8 umol/l.Sample c
on (b)(6) 2026:
results from the c 503 analyzer:
the initial result was 1.21 umol/l.The repeat result in reduced sample volume was 16.8 umol/l.This patient sample was rerun on a cobas pure analyzer:
the repeat results of a manual 3-fold dilution and an automated 5-fold dilution were approximately 96.3 µmol/l.Repeat results of 32.1 umol/l and 19.1 umol/l were also provided.On (b)(6) 2026:
the repeat result at a dilution ratio of 50 was >60 umol/l.The repeat result at a manual dilution ratio of 50 was 92 umol/l.Sample d
the result at a manual dilution ratio of 50 was 63.3 umol/l.Patient 2
sample a
on (b)(6) 2026:
the initial result was 1.21 umol/l with a data flag.The repeat result at reduced sample volume was 16.8 umol/l.The repeat result was 58.6 umol/l with a data flag.Sample b
on (b)(6) 2026:
the initial result was 1.20 umol/l.The first repeat result was 1.20 umol/l with a data flag.The second repeat result in reduced sample volume was 16.7 umol/l.On (b)(6) 2026:
the third repeat result was 1.2 umol/l
the fourth repeat result at a dilution ratio of 50 was 58.9 umol/l with a data flag.--
the fifth repeat result at a dilution ratio of 150 was 75.9 umol/l.Sample c
the result at a dilution ratio of 50 was >60 umol/l.Patient 3
sample a
on (b)(6) 2026:
the initial result was 1.21 umol/l with a data flag.The repeat result at reduced sample volume was 16.8 umol/l.Sample b
on (b)(6) 2026:
the initial result was 1.18 umol/l.The first repeat result in reduced sample volume was 16 umol/l.The second repeat result in reduced sample volume was 16 umol/l.The third repeat result at a dilution ratio of 50 was 15.7 umol/l with a data flag.
|
| |
|
Additional Manufacturer Narrative
|
|
The cobas c 503 analytical unit serial number is (b)(6).The cobas pure c 303 analyzer serial number and the online tdm methotrexate reagent lot number used for this analyzer were not provided.
|
| |
|