It was reported to boston scientific corporation that a speedband superview super 7 device was used in the esophagus during an esophageal variceal ligation procedure performed on (b)(6) 2026 for the treatment of esophageal variceal ligation (evl).During the procedure, when the third band was deployed, it was found that the bands were twisted together and it could not be deployed.There was no difficulties setting up the device.The procedure was completed with another of the same device.There were no patient complications reported as a result of this event.The patient's condition at the conclusion of the procedure was reported to be stable.
|
|
Block h6:
imdrf device code a050501 captures the reportable event of bands failure to deploy.Block e1: initial reporter facility name = (b)(6).Block h11: investigation results
the speed band super view super 7 device was not returned for analysis; therefore, a technical analysis could not be performed.A review of the device history record (dhr) was performed.It was confirmed this device met manufacturing specification prior to distribution and there were no manufacturing deviations which could have contributed to the reported event.A similar complaint review was completed.There were no emerging trends identified that warrant further action.Based on the event description, information and media provided by the customer there is not enough evidence to determine whether the issue was induced due to the physicians technique during the procedure or were related to a device malfunction.Additionally, based on the most relevant information for the complaint including product record review results, the device meets all manufacturing specifications required and passed all the controls and inspections, no abnormalities were reported during the assembly process, the definitive cause of the reported issue cannot be established due to lack of evidence.Without proper evaluation of the device, it remains unknown the causes that contributed to the event.Based on the thorough review of the reported complaint, boston scientific has assigned an investigation conclusion code of cause not established.
|