It was reported that an inaccuracy between the continuous glucose monitor (cgm) and the blood glucose (bg) meter occurred.On 08/09/2026, the patient experienced lethargy and increased urination.The patient was brought to the hospital and when a fingerstick was taken the cgm reading was low, and the blood glucose reading was 500 mg/dl.Ketones were high, but no specific value was reported.The patient experienced diabetic ketoacidosis and was treated with intravenous insulin and fluids.No further medication was reported and the patient was discharged after spending less than 24 hours at the hospital.There was no documentation of further medical evaluation or intervention.At the time of the report the patient was stable.Data was evaluated and the allegation was undetermined.The probable cause could not be determined via data.The reported glucose values fall within the d zone of the parkes error grid.The data investigation did not find blood glucose calibration values to compare to cgm values during the reported sensor session or the calibrations during the reported sensor session were in accurate range per device specifications.No additional patient or event information is available.
|