Sales rep reported on (b)(6) 2026 that a when a surgeon was implanting a vbeal-3-7, a tpeg2.4 screw passed through the plate.The surgeon had to remove the already implanted screws and plate to switch to a new vbeal-3-7s to complete the surgery which took approx.30 minutes.The second plate was implanted successfully with no further issues.The rep was not present during the case but after they spoke to the r4, he believed the bearing was disrupted when they drilled using the single sided tissue protector.After trimed product management spoked with the rep, the rep is unsure if the bearing popped out and did not know which screw hole was affected.Another sales rep stated on (b)(6) 2026 that the surgeon did not use the term that the bearing popped out but when they advanced the screw the screw basically broke through the bearing using the tpeg 2.4.
|