A steris service technician arrived onsite following the reported event to inspect the transfer carriage manifold and found that the front wheels did not properly engage after the manifold was pulled out of the sterilizer, causing the manifold to fall to the floor.After further inspection, the technician found that the leg lock assembly and release mechanism were misaligned, preventing proper operation of the front wheel locking system.To resolve the issue, the technician adjusted and realigned the front assembly of the transfer carriage, including the leg lock assembly and release mechanism.After the equipment was tested and confirmed operating according to specifications, it was returned to service.No additional issues have been reported.
|