| Model Number |
RLV 0065 |
| Medical Device Problem Code |
Material Integrity Problem (2978)
|
| Health Effect - Clinical Code |
Foreign Body In Patient (2687)
|
| Date of Event |
08/19/2026
|
|
Type of Reportable Event
|
Serious Injury
|
|
Event or Problem Description
|
|
It was reported that during a basivertebral nerve (bvn) radiofrequency ablation (rfa) procedure, the bevel introducer cannula fractured while being advanced into the first vertebral body.The physician was malleting the introducer cannula when the clip on the bevel introducer cannula broke.It is unknown whether any fragmented material remained within the patient or if it was contained within the vertebral body.An additional device was used to complete the procedure.The patient is doing well, and no adverse effects were reported.The device is expected to be returned for analysis.
|
| |
|
Event or Problem Description
|
|
It was reported that during a basivertebral nerve (bvn) radiofrequency ablation (rfa) procedure, the bevel introducer cannula fractured while being advanced into the first vertebral body.The physician was malleting the introducer cannula when the clip on the bevel introducer cannula broke.It is unknown whether any fragmented material remained within the patient or if it was contained within the vertebral body.An additional device was used to complete the procedure.The patient is doing well, and no adverse effects were reported.The device is expected to be returned for analysis.\
\
additional information was received stating that the fractured component was completely removed.The fracture was located at the top of the t handle body.No fragmented material remained within the patient.
|
| |
|