B2 outcomes attrib to adv event: corrected.A4 weight: 82.5 kg.E1 initial reporter address 1: (b)(6).E1 initial reporter phone: (b)(6).With all the available information, boston scientific concludes:
device analysis findings:
the device was not returned for analysis; therefore, a technical analysis could not be performed.Device history record (dhr) review:
review confirmed that the device met all material, assembly, and performance specifications prior to distribution.Labeling review:
review of the instructions for use (ifu) confirmed there was relevant content and sufficient guidance with respect to the circumstances described within this complaint.No updates are required to the ifu as a result of this event.Risk review:
a risk review was performed and confirmed that the event was defined in the risk documentation.This event type has been accounted for during product risk analysis to support acceptable risk benefit for the product.Investigation conclusion:
based on a thorough review of the reported complaint, the most probable cause for this complaint was considered adverse event related to procedure.The instructions for use include testing of the device before usage within the body and state that use of the device is to be discontinued if evidence of a failure or damage is present.As the reported events do not indicate any device damages or failures during unpackaging, preparation, or testing, it was considered likely that the device events and following patient events occurred due to factors encountered during the procedure.
|