| Model Number |
SOLOFUSE1 |
| Medical Device Problem Code |
Intermittent Capture (1080)
|
| Health Effect - Clinical Code |
Aortic dissection (4870)
|
| Date of Event |
07/02/2026
|
|
Type of Reportable Event
|
Serious Injury
|
|
Event or Problem Description
|
|
The following was reported: "customer reports they were completing a tavr with medtronic evolut valve.Shallow core valve and embolized on post dil.Patient did not need open surgery, they left the valve in place and placed a new evolut medtronic valve."
the valve embolization resulted in a vessel dissection that was addressed through vascular surgical intervention.The patient did not require any additional treatment following the intervention, and the patient outcome was reported as good.
|
| |
|
Additional Manufacturer Narrative
|
|
During a follow-up interview with the user, it was determined that loss of capture occurred during the procedure.Review of the procedural imaging identified that the fusion wire was positioned on top of the mitral valve apparatus where loss of contact is a known risk during valve deployment when balloon inflation shifts the relationship of the conducting part of the lead wire to the myocardium and leads to loss of contact and resultant loss of capture.A review of the epg logs did not identify any pacing or sensing issues.A review of the production records did not identify an issue with the epg or fusion components.Based on the information obtained from the customer and review of the imaging, potential contributing factors to the valve embolization included the shallow positioning of the transcatheter valve and the position of the fusion wire, which resulted in loss of capture.
|
| |
|