• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: Lumenis Ltd Lumenis Moses 365 D\F\L; powered laser surgical instrument

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

Lumenis Ltd Lumenis Moses 365 D\F\L; powered laser surgical instrument Back to Search Results
Model Number 1912-10
Medical Device Problem Codes Break (1069); Thermal Decomposition of Device (1071)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 08/19/2026
Type of Reportable Event Malfunction
Event or Problem Description
It was reported that during an ureteroscopy procedure to treat kidney stone, the fiber was plugged in, and the surgeon was firing on the stone when the tip of the fiber broke off.The procedure was completed using another fiber of the same model.There was no patient complication.
 
Event or Problem Description
It was reported that during an ureteroscopy procedure to treat kidney stone, the fiber was plugged in, and the surgeon was firing on the stone when the tip of the fiber broke off.The procedure was completed using another fiber of the same model.There was no patient complication.
 
Additional Manufacturer Narrative
The device is not available for analysis; therefore, no physical analysis of the product could be performed.The reported device performance allegation cannot be confirmed.A device history record (dhr) review indicates the device met all manufacturing specifications, and therefore the failure was unlikely related to manufacturing.The labeling review found that the product instruction for use (ifu) states: inspect the fiber for kinks, punctures, fractures, or other damage.If the fiber appears damaged, do not use the fiber; return it to the supplier for replacement and hold the fiber tip to a non-reflective surface, and ensure that a circular red/green spot appears.If the spot is not circular, cleave the fiber.If the spot is weak or not visible, discard the fiber or return it to the supplier for replacement.A risk review was completed and confirmed that the event of fiber cap/tip - blew/burned out and fiber cap/tip - break are defined in the risk documentation.Based on the information available the cause that contributed to the reported event cannot be established due to lack of evidence.Since the investigation findings do not lead to a clear conclusion about the cause of the reported event, the investigation conclusion code selected for this complaint is cause not established.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
Lumenis Moses 365 D\F\L
Common Device Name
powered laser surgical instrument
Manufacturer (Section D)
Lumenis Ltd
hakidma 6
yokneam 20692 04
IS  2069204
Manufacturer (Section G)
ADVANCED FIBER TOOLS GMBH
bornheimer str 4
mittweida 09468
GM   09468
Manufacturer Contact
farshad fahimi
4100 hamline avenue north
building c
saint paul, MN 55112
MDR Report Key50699658
Report Number2124215-2026-48293
Device Sequence Number14925432
Product Code GEX
UDI-Device Identifier07290109143408
UDI-Public(01)07290109143408(17)280723(10)18222512
Combination Product (Y/N)N
Initial Reporter StateOH
Initial Reporter CountryUS
PMA/510(K) Number
K170121
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source Health Professional
Initial Reporter Occupation Physician
Type of Report Initial
Report Date (Section B) 09/09/2026
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model Number1912-10
Device Catalogue Number1912-10
Device Lot Number0018222512
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 08/19/2026
Initial Report FDA Received Date09/09/2026
Date Device Manufactured01/22/2026
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient SexUnknown
-
-