Shockwave Medical, Inc. Shockwave IVL System with the Shockwave C2 Aero Coronary IVL Catheter
|
Back to Search Results |
|
| Model Number |
C2AIVL2512 |
| Medical Device Problem Code |
Material Split, Cut or Torn (4008)
|
| Health Effect - Clinical Code |
Perforation of Vessels (2135)
|
| Date of Event |
08/14/2026
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
The device referenced in this report was not returned to shockwave medical, inc.Therefore, a physical examination cannot be performed.The cause of the perforation could not be definitively determined based on the information available.\
\
balloon ruptures are a known failure mode with a low probability of occurrence.Contributing factors include patient calcium, damage caused when the balloon wall comes into close contact with the device emitters, likely causes include (a) an irregular calcium lesion capable of compressing the balloon wall into close proximity with the emitter(s) and/or (b) an incompletely inflated balloon.\
\
the shockwave intravascular lithotripsy (ivl) system with the shockwave c2 aero coronary ivl catheter generates acoustic pressure pulses within the target treatment site, disrupting calcium within the lesion allowing subsequent dilatation of a coronary artery stenosis using low balloon pressure not to exceed 2-4 atm during ivl treatment, and 6 atm for post dilatation.The pressure used for the ivl balloon catheter is much lower than other balloon catheters used in percutaneous coronary intervention (pci), thus the risk associated with loss of balloon pressure during ivl catheter use is negligible.The associated patient risk for this failure mode is low and adequately captured in swmi's risk documentation.\
\
shockwave medical has controls in place to ensure devices are built to approved procedures and meet lot release acceptance criteria prior to being distributed.A review of the manufacturing and test documentation for subject lot does not reveal any issues with the manufacturing of the device.The device passed all of shockwave medical, inc.Acceptance criteria prior to shipping.
|
| |
|
Event or Problem Description
|
|
Two shockwave c2 aero coronary intravascular lithotripsy (ivl) catheters were utilized during treatment of lesions located in the obtuse marginal (om) artery and mid circumflex artery.\
\
the om artery lesion was initially pre-dilated using a trek¿ 2.5 × 15 mm dilatation catheter.A shockwave c2 aero 2.5 × 12 mm ivl catheter was subsequently advanced to the target lesion and delivered 70 ivl pulses at inflation pressures of 4 atm and 6 atm.Following ivl therapy, a stent was successfully placed and post-dilatation was performed using a non-compliant (nc) balloon.No device-related malfunctions were reported with the first ivl catheter.\
\
next, the lesion in the mid circumflex artery was pre-dilated using a trek¿ 2.5 × 15 mm dilatation catheter.A second shockwave c2 aero 2.5 × 12 mm ivl catheter was advanced to the target lesion and delivered 10 ivl pulses at 4 atm, followed by an additional 20 ivl pulses at 6 atm.During ivl therapy, the balloon reportedly burst.Concurrently, a vessel perforation was observed.The reported perforation was managed with balloon tamponade.The patient was reported to be recovering, with no further adverse clinical sequelae or complications reported.
|
| |
|