Baxter received a report that operating table trusystem 7000 u (dv) had table will move without any control button press.The process step during which this occurred was unknown.There was no patient/user injury, medical intervention, symptom, or adverse event reported.
|
|
The baxter technician found key membrane error.The trusystem 7000 operating table is intended for the following use:
¿ patient positioning during surgery, including anaesthesia induction and recovery from anaesthesia.¿ task-related patient transfer within the operating theatre.¿ for applications in conjunction with intra-operative imaging (e.G.Ct, mrt, x-ray) with specific table components.The technician replaced the control panel to resolve the reported event.Based on this information, no further action is required.Although there was no reported injury with this event, if the report of unintended movement were to recur, it could potentially cause serious injury or death.Therefore, baxter is reporting this event.
|