BOSTON SCIENTIFIC CORPORATION ACUITY? X4 Straight; Implantable Pulse Generator, Pacemaker (Non-Crt) Back to Search Results
Model Number 4672
Medical Device Problem Codes Failure to Capture (1081); High Capture Threshold (3266)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 08/10/2026
Type of Reportable Event Malfunction
Event or Problem Description
It was reported that this left ventricular (lv) lead had high thresholds and lv loss of capture (loc) was observed on the presenting electrogram (egm).The programmed lv output was increased, and lv revision may be considered at battery replacement time.This lv lead remains in service.No adverse patient effects were reported.
 
Additional Manufacturer Narrative
The local area field representative was contacted for additional information regarding event cause and resolution.At this time, no further information is available.Should additional information become available this report will be updated.This product investigation is still in progress.This report will be updated when product investigation is complete.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ACUITY? X4 Straight
Common Device Name
Implantable Pulse Generator, Pacemaker (Non-Crt)
Manufacturer (Section D)
BOSTON SCIENTIFIC CORPORATION
4100 hamline avenue north
saint paul MN 55112
Manufacturer (Section G)
Boston Scientific Corporation
4100 hamline avenue north
saint paul MN 55112 -579
Manufacturer Contact
mark mischke
4100 hamline avenue north
saint paul, MN 55112
+1(612)875-7234
MDR Report Key50704037
Report Number2124215-2026-48357
Device Sequence Number9614596
Product Code LWP
UDI-Device Identifier00802526524530
UDI-Public(01)00802526524530(17)170705
Combination Product (Y/N)Y
Initial Reporter StateNY
Initial Reporter CountryUS
PMA/510(K) Number
P010012
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2017
Reporter Type Manufacturer
Report Source Health Professional
Initial Reporter Occupation Physician Assistant
Type of Report Initial
Report Date (Section B) 09/09/2026
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Lay User/Patient
Device Expiration Date07/05/2017
Device Model Number4672
Device Catalogue Number4672
Device Lot Number502447
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 08/11/2026
Initial Report FDA Received Date09/10/2026
Date Device Manufactured01/05/2016
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient Age78 YR
Patient SexMale
-
-