• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: Tosoh Hi-Tec, Inc. HLC-723 G8; Hemoglobin A1c Test System

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

Tosoh Hi-Tec, Inc. HLC-723 G8; Hemoglobin A1c Test System Back to Search Results
Model Number HLC-723G8 ANALYZER
Medical Device Problem Code Incorrect, Inadequate or Imprecise Result or Readings (1535)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 08/17/2026
Type of Reportable Event Malfunction
Event or Problem Description
A customer reported failed gh5-b 2026 college of american pathologists (cap) survey proficiency tests on the hlc-723 g8 analyzer.Value gh-09 failed and the remaining four values were high, but within range.The tosoh quality assurance (qa) laboratory passed all cap gh5-b 2026 proficiency tests.The customer requested service.Field service was dispatched.There was no indication of any patient intervention or adverse health consequences due to the reported event.
 
Additional Manufacturer Narrative
The most probable cause is not yet determined, investigation pending.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
HLC-723 G8
Common Device Name
Hemoglobin A1c Test System
Manufacturer (Section D)
Tosoh Hi-Tec, Inc.
1-37, fukugawa minami-machi
shunan, yamaguchi 74600 42
JA  7460042
Manufacturer (Section G)
Tosoh Hi-Tec, Inc.
1-37, fukugawa minami-machi
shunan, yamaguchi 74600 42
JA   7460042
Manufacturer Contact
janefrances ajiduah
2-2-1 yaesu
chuo-ku, tokyo 10400-28
JA   1040028
+1(650)636-8314
MDR Report Key50704489
Report Number3004529019-2026-00125
Device Sequence Number20327453
Product Code PDJ
UDI-Device Identifier04560189282919
UDI-Public04560189282919
Combination Product (Y/N)N
Initial Reporter StateNJ
Initial Reporter CountryUS
PMA/510(K) Number
K200904
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Health Professional
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial
Report Date (Section B) 09/09/2026
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model NumberHLC-723G8 ANALYZER
Device Catalogue Number021560
Device Lot NumberN/A
Was Device Available for Evaluation? Yes
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 08/17/2026
Initial Report FDA Received Date09/10/2026
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured04/01/2021
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Reuse
Patient Sequence Number1
Patient SexUnknown
-
-