| Model Number |
NIM4CM01 |
| Medical Device Problem Codes |
False Negative Result (1225); Incorrect, Inadequate or Imprecise Result or Readings (1535)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
08/18/2026
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
It was reported that intra-op, error code was received after the start, the console had an alarm problem 'out of the measurement range' messages during surgery, which were resolved by checking the electrodes.The second more important problem, when approaching the nerve, the beeping sound was present but no feedback on the curve on the screen during the intervention despite the verification of consumables and connections, there was no fault with the other components.There was no patient impact.
|
| |
|
Additional Manufacturer Narrative
|
|
H6: the codes fdm b16, fdr c20, fdc d16 and img g02030 are applicable for nim console (nim4cm01).Continuation of d10: section d information references the main component of the system.Other relevant device(s) are: product id: nim4cpb1, serial/lot #: p2214239, ubd: , udi#: 00763000395902 h6: the codes fdm b16, fdr c20, fdc d16 and img g02005 are applicable for nim patient interface(nim4cpb1).Medtronic submits this report to comply with fda regulations 21 cfr parts 4 and 803.Medtronic has made reasonable efforts to provide as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.Medtronic will submit a supplemental report if additional relevant information becomes known.
|
| |
|
Search Alerts/Recalls
|
|
|