| Model Number |
D150 |
| Medical Device Problem Code |
Over-Sensing (1438)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
08/28/2026
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
It was reported that the healthcare professional (hcp) for the patient with this implantable cardioverter defibrillator (icd) contacted technical services (ts) regarding a large signal observed during a ventricular episode.Ts reviewed the available data and noted that the signal appeared consistent with cardiac contractility modulation (ccm) therapy, which would not be unexpected if a ccm system were implanted.The hcp was unable to confirm whether the patient had a ccm device.Ts further noted that the signal was intermittently oversensed; however, it occurred within the device's blanking period and therefore did not affect timing cycles or result in a device response.This device remains in service.No adverse patient effects were reported.
|
| |
|
Additional Manufacturer Narrative
|
|
If pertinent information is provided in the future, a supplemental report will be submitted.
|
| |
|