• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: Datascope Corp. CARDIOSAVE HYBRID TYPE D PLUG; System, Balloon, Intra-Aortic And Control

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

Datascope Corp. CARDIOSAVE HYBRID TYPE D PLUG; System, Balloon, Intra-Aortic And Control Back to Search Results
Model Number 0998-00-0800-33
Medical Device Problem Codes Charging Problem (2892); Power Problem (3010)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 08/21/2026
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Other contact phone number: (b)(6)\ updated fields: b1, b4, g3, g6, h2, h6(type of investigation, investigation findings, investigation conclusions), h11 \ correction fields: d10, e1(event site name), e2, e3, g2, g7, h6(medical device problem code)\ revert a1,h4 should left blank\ a getinge field service engineer (fse) was dispatched to evaluate the unit.The fse reported that upon checking the machine the mentioned issue was observed.The fse checked the machine and found no mains voltage on the power cord and power supply.The fse reported that the device was not showing the mains indicator and not charging the batteries as well.The fse mentioned that the reel ac power cord required replacement.After replacement the fse found that the mentioned failure was resolved.The fault logs were requested for review but were not provided by the customer.The unit passed all safety and functional checks successfully.The device was returned to the customer and cleared for clinical use.\ based on fse investigation the effect of power supply failure or power cord or plug stuck or unable to be pulled out or no ac power or damaged on the cardiosave is confirmed.
 
Event or Problem Description
N/a.
 
Additional Manufacturer Narrative
Updated fields: b4, g3, g6, h2, h11\ corrected field- g2.
 
Event or Problem Description
N/a.
 
Additional Manufacturer Narrative
Event site postal code: (b)(6).\ upon completion of the investigation into this event a follow up report will be submitted.
 
Event or Problem Description
It was reported by the fse that the cardiosave intra-aortic balloon pump (iabp) is not showing main supply.There was no patient involved.No harm was reported.A getinge field service engineer (fse) was dispatched to evaluate the unit.The fse reported that the machine was checked and no mains voltages were detected on the power cord or power supply.Unit is not showing the mains supply indicator and not charging the batteries as well.Fse found that below mentioned spare are defective.Replaced the same and resolved the issue.\ device passed the all safety and functional checks successfully.Device was returned to customer and cleared for clinical use.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CARDIOSAVE HYBRID TYPE D PLUG
Common Device Name
System, Balloon, Intra-Aortic And Control
Manufacturer (Section D)
Datascope Corp.
1300 macarthur blvd
mahwah NJ 07430
Manufacturer (Section G)
Datascope Corp.
1300 macarthur blvd
mahwah NJ 07430
Manufacturer Contact
ewelina chorlowska
1300 macarthur blvd
mahwah, NJ 07430
MDR Report Key50708896
Report Number2249723-2026-0004234
Device Sequence Number9852722
Product Code DSP
UDI-Device Identifier10607567109008
Combination Product (Y/N)N
Initial Reporter CountryIN
PMA/510(K) Number
K112372
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source Foreign,Company Representative
Initial Reporter Occupation Biomedical Engineer
Type of Report Initial
Report Date (Section B) 09/10/2026
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model Number0998-00-0800-33
Device Catalogue Number0998-00-0800-33
Was Device Available for Evaluation? Yes
Is the Reporter a Health Professional? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 08/21/2026
Initial Report FDA Received Date09/10/2026
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured10/03/2024
Is the Device Labeled for Single Use? (Y/N) No
Usage of Device Reuse
Patient Sequence Number1
Patient SexUnknown
-
-