| Model Number |
LifeVest WCD 4000 System |
| Medical Device Problem Code |
Charging Problem (2892)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
08/26/2026
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
A us distributor reported that a charger was unable to power on.
|
| |
|
Additional Manufacturer Narrative
|
|
Device evaluation of charger has been completed.The reported problem (will not power on) was confirmed due to a damaged power supply connector.The root cause for the damaged power supply connector was unable to be positively identified.There was no adverse event that resulted from the damaged charger.
|
| |
|