| Catalog Number |
NNU8LPT |
| Medical Device Problem Code |
Fluid/Blood Leak (1250)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
08/19/2026
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
On (b)(6), 2026, a patient underwent an ultrasound guided ascites percutaneous drainage catheter placement procedure using a navarre universal drainage catheter.During the procedure, it was reported that leakage was observed from the drainage catheter.The procedure was subsequently completed using another device.There was no reported patient injury.
|
| |
|
Event or Problem Description
|
|
On (b)(6) 2026, a patient underwent an ultrasound guided ascites percutaneous drainage catheter placement procedure using a navarre universal drainage catheter.During the procedure, it was reported that leakage was observed from the drainage catheter.The procedure was subsequently completed using another device.There was no reported patient injury.
|
| |
|
Additional Manufacturer Narrative
|
|
H11: as the lot number for the device was provided, a review of the device history records was performed.The device has not been returned to the manufacturer for evaluation.However, photos were provided for review.The investigation of the reported event is currently underway.Section a through f: the information provide by bd represents all the known information at this time.Despite good faith efforts to obtain additional information, the complainant/reporter was unable or unwilling to provide any further patient, product, or procedural details to bd.
|
| |
|
Additional Manufacturer Narrative
|
|
H11: manufacturing review: the device history records have been reviewed, and this lot met all release criteria.Investigation summary: although the physical device was not returned for evaluation, three electronic photos were provided for review.The first photo shows the partial view of a drainage catheter and thread was noted near the catheter.The second photo shows the partial view of a drainage catheter, and the distal end of the catheter was noted to be in pigtail formation.The thread was also noted near the distal end of catheter.The third photo shows the complete view of a fully loaded navarre drainage catheter, in which thread was noted be separated since it is not reported considering it as incidental.No visible leak can be seen from all the three devices.The investigation is inconclusive for reported fluid leak issue for all the three devices as the provided evidence was not sufficient to confirm the issue.The definitive root cause could not be determined based upon available information.Labeling review: as the reported event did not allege a labeling or use related issue, a labeling review is not required.G3, h6 (method, result, conclusion)
e1 (b)(6) hospital
section a through f: the information provide by bd represents all the known information at this time.Despite good faith efforts to obtain additional information, the complainant/reporter was unable or unwilling to provide any further patient, product, or procedural details to bd.
|
| |
|
Search Alerts/Recalls
|