MAUDE Adverse Event Report: BOSTON SCIENTIFIC CORPORATION Percuflex Plus; catheter, biliary, diagnostic

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BOSTON SCIENTIFIC CORPORATION Percuflex Plus; catheter, biliary, diagnostic Back to Search Results
Model Number M0061752530
Medical Device Problem Code Break (1069)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 06/02/2026
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Block h6:\ imdrf device code a0401 captures the reportable event of stent shaft break.\ \ block h11:\ the device was not returned for analysis; however, the customer provided a photo.During inspection, it was observed that the shaft was detached.It is possible to conclude that this issue could be caused by operational factors.If the retrieval line gets entangled in the shaft and is removed using excess force, the stent can get detached or separated during the preparation affecting the performance of the device.A review of the device history record (dhr) was performed.It was confirmed this device met manufacturing specification prior to distribution and there were no manufacturing deviations which could have contributed to the reported event.A risk review was completed and confirmed that the event of stent shaft break was defined in the risk documentation and is documented accordingly.This event type has been accounted during product risk analysis to support acceptable risk benefits for the product.Therefore, all compiled information on this investigation determines that the most probable cause is adverse event related to procedure since the adverse event occurred during the procedure and the device had no influence on the event.
 
Event or Problem Description
It was reported that a percuflex plus ureteral stent was used during transurethral laser lithotripsy and stone removal of the ureter/renal pelvis plus transurethral ureteral stent placement under general anesthesia with endotracheal intubation.During procedure, it was discovered that the stent was broken.The procedure was completed with another of the same stent.There were no patient complications as a result of this event.
 
Event or Problem Description
It was reported that a percuflex plus ureteral stent was used during transurethral laser lithotripsy and stone removal of the ureter/renal pelvis plus transurethral ureteral stent placement under general anesthesia with endotracheal intubation.During procedure, it was discovered that the stent was broken.The procedure was completed with another of the same stent.There were no patient complications as a result of this event.
 
Additional Manufacturer Narrative
Block h6:\ imdrf device code a0401 captures the reportable event of stent shaft break.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
Percuflex Plus
Common Device Name
catheter, biliary, diagnostic
Manufacturer (Section D)
BOSTON SCIENTIFIC CORPORATION
300 boston scientific way
marlborough MA 01752
Manufacturer (Section G)
Boston Scientific Corporation
2546 calle primera
alajuela
CS  
Manufacturer Contact
farshad fahimi
4100 hamline avenue north
building c
saint paul, MN 55112
MDR Report Key50713512
Report Number2124215-2026-48398
Device Sequence Number18656186
Product Code FAD
UDI-Device Identifier08714729056812
UDI-Public(01)08714729056812(17)280910(10)37432155
Combination Product (Y/N)N
Initial Reporter CountryCH
PMA/510(K) Number
K924608
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source Other,Foreign,Health Professional,Company Representative
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial
Report Date (Section B) 09/10/2026
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model NumberM0061752530
Device Catalogue Number175-253
Device Lot Number0037432155
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 08/15/2026
Initial Report FDA Received Date09/10/2026
Date Device Manufactured09/11/2025
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient Age60 YR
Patient SexMale
-
-