| Model Number |
P3200C |
| Medical Device Problem Code |
Unintended Movement (3026)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
08/17/2026
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
Baxter received a report that versacare frame had brakes not holding while in brake position.The process step during which this occurred was unknown.There was no patient/user injury, medical intervention, symptom, or adverse event reported.
|
| |
|
Additional Manufacturer Narrative
|
|
The baxter technician found the casters needed to be replaced.A search of the baxter maintenance records showed baxter performed preventative maintenance on this bed in dec 10, 2025.It is unknown if the facility performed any other preventative maintenance on this bed.Per the baxter service manual, perform annual preventive maintenance procedures to make sure all versacare bed components are functioning as originally designed.Brake casters should be checked for cuts, wear and quality of tread, etc.And replaced when necessary.Examine the brakes to see whether the bed moves when the brakes are set.Replace as necessary.The technician replaced the casters to resolve the reported event.Based on this information, no further action is required.
|
| |
|