It was reported that split drape with adhesive (b)(6) no longer sticks appropriately and this has been continuing for months.Medline reviewed the complaint history and nonconformance trends for this product and this customer for the past 24 months; there are no additional complaints for this customer and one additional unrelated complaint for this product.One each dynjp2498 was received and the sample was reviewed; the adhesive adhered to the skin with no issue, and the manufacturer is using the appropriate adhesive.Based on the returned sample and the product trending the issue is not confirmed; the cause was listed as not confirmed with received sample, and based on the investigation results, no further actions are being taken at this time.There were no reports of death, serious injury, medical intervention, or follow up care.It has been determined that the reported event could cause or contribute to serious injury if it were to occur again.In an abundance of caution, this is a reportable event.If additional information becomes available this report will be reopened and reevaluated.This medical device report (mdr) is being submitted as part of retrospective review activities, which include review of historical data in accordance with regulations at 21 cfr part 803.The information included in this mdr reflects the information reasonably known to the manufacturer at the time of this retrospective review and submission.
|