A steris service technician arrived onsite following the reported event to inspect the transfer cart and found that would not consistently latch in the latching station at the bottom of the sterilizer.After further inspection, the technician found that the latching mechanism was out of adjustment, causing intermittent latching.To resolve the issue, the technician adjusted the latching mechanism.After the unit was tested and confirmed operating according to specifications it was returned to service.This unit is not under steris contract agreement for maintenance services.The user facility is responsible for all maintenance activities.No additional issues have been reported.
|