(b)(4).Date sent 9/30/2026.D4 batch # 141e95.Corrected data d4: udi#.Investigation summary:
the product was returned for thorough evaluation, which included visual inspections and functional testing of the returned device.Visual analysis of the returned sample revealed that the cdh29p device arrived with no apparent damage.The breakaway washer was present, cut with the halves located in the anvil area and there were no staples present.When the test battery was installed the green checkmark illuminated, indicating that the device was fired and achieved complete firing stroke.The device was reloaded with staples, a new washer was placed on the device.The device was reset and tested for functionality.It fired and formed all the staples as well as completely cut the test media and the breakaway washer without incident.The staple line was complete, and the staples meet the staple form release criteria.The event described could not be confirmed as the device was returned without detectable damage.Although no product defect was identified, there may have been other circumstances or issues that occurred during the use of the device that could not be replicated during the laboratory analysis.As part of ethicon endo surgery¿s quality process, all devices are manufactured, inspected, and released to approved specifications.A manufacturing record evaluation was performed for the finished device batch number 141e95, and no non-conformances were identified.
|