It was reported a tha revision was performed for a disassociated liner after the patient had fell down while playing pickle ball.The patient was about 3 year post-op on the day of the revision.After exposure and dislocation of the hip, and removal of the femoral head, the surgeon inspected the liner.The liner (altrx 36/52 +4 / 10 degrees) was not really disassociated but rather had broken into a few pieces.The elevated portion of the liner (the 10 degree lip) had broken off, as well as some of the ards.The stem in situ is a corail kla size 13.The femoral head had been rubbing directly against the cup for some time, so there was some intra-articular debris.The surgeon opted to remove the cup and upsize to a size 54.He used an emphasys multi-hole and elevated rim liner.He then used a ts delta ceramic head size 40+5.The corail stem was well fixed and remained in situ.The post x-rays were satisfactory and the patient is in good condition.
|
|
Product complaint # (b)(4).This report is being submitted pursuant to the provisions of 21 cfr, part 803 (and/or part 4, as applicable).This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by depuy synthes, or its employees that the report constitutes an admission that the product, depuy synthes, or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
|