• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: Medela AG Freestyle Hands-free EN ES FR; Pump, Breast, Powered

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

Medela AG Freestyle Hands-free EN ES FR; Pump, Breast, Powered Back to Search Results
Model Number 101044164
Medical Device Problem Code Suction Problem (2170)
Health Effect - Clinical Code Inflammation (1932)
Date of Event 07/06/2026
Type of Reportable Event Serious Injury
Event or Problem Description
On (b)(6) 2026, the customer contacted medela llc that while using the freestyle hands-free she was diagnosed with mastitis due to poor or insufficient suction.Additionally, on (b)(6) 2026 she alleged that she was prescribed antibiotics.
 
Additional Manufacturer Narrative
Customer service sent a replacement pump to the customer and asked for the return of her old pump for evaluation.In follow up with a complaint handler on (b)(6) 2026, the customer stated that she was sent a replacement pump and she had sent her pump in for evaluation.Additionally, the customer stated she had developed mastitis requiring antibiotics approximately one month before contacting medela.On (b)(6) 2026, the pump was evaluated and tested within normal operating range the customers complaint was unable to be duplicated.Based on the results of our internal investigation (capa-2025-0027) including published literature, it cannot definitively be concluded that the pump caused or has contributed to the customer's reported mastitis.According to published clinical literature, mastitis incidence in lactating women, with or without a breast pump, can be as high as 33%.In fact, clinical guidelines recommend the use of a breast pump to facilitate the removal of breast milk during periods of mastitis.Additionally, complaint trends across all medela pumps show mastitis incidence consistently below 0.1%, far lower than the 3-20% occurrence in the general population of lactating women "[2]".
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
Freestyle Hands-free EN ES FR
Common Device Name
Pump, Breast, Powered
Manufacturer (Section D)
Medela AG
4b lattichstrasse
baar zug 6340
SZ  6340
Manufacturer (Section G)
Medela AG
lattichstrasse 4b
baar zug 6340
SZ   6340
Manufacturer Contact
alea raber
1101 corporate drive
mchenry, IL 60050
+1(800)435-8316
MDR Report Key50714764
Report Number1419937-2026-00028
Device Sequence Number10598769
Product Code HGX
UDI-Device Identifier00020451441641
UDI-Public00020451441641
Combination Product (Y/N)N
Initial Reporter CountryCA
PMA/510(K) Number
K191653
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source Consumer
Type of Report Initial
Report Date (Section B) 09/10/2026
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Lay User/Patient
Device Model Number101044164
Device Catalogue Number101044164
Device Lot Number0000753138
Was Device Available for Evaluation? Yes
Date Returned to Manufacturer08/25/2026
Is the Reporter a Health Professional? No
Initial Date Received by Manufacturer 08/18/2026
Initial Report FDA Received Date09/10/2026
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured01/27/2026
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Reuse
Patient Sequence Number1
Outcome Attributed to Adverse Event Other;
Patient SexFemale
-
-