A device history record, complaint history review, and risk review on failure modes are performed on each device with reportable malfunction(s) along with other methods of investigation as coded in section h6 of this mdr report.Investigation summary:
field technician stated issue of failed leak down test was not confirmed.On 16-sep-26, etco2 was received unboxed and with paperwork.Passed asm leak down test.No faults found.Device to be returned for processing.Unit was opened and calibration required.Failure investigation report attached to sales force.This report lists imdrf annex a, c, d, and g codes that are reportable malfunctions observed on the device during servicing.Per 803.52(f)(11)(iii) the information provided was obtained from servicing activities performed on the device.There were no additional details obtainable or provided at the time of service.
|