MAUDE Adverse Event Report: BOSTON SCIENTIFIC CORPORATION AMS 700 LGX MS Pump; prosthesis, penis, inflatable

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BOSTON SCIENTIFIC CORPORATION AMS 700 LGX MS Pump; prosthesis, penis, inflatable Back to Search Results
Model Number 72404252-10
Medical Device Problem Codes Degraded (1153); Material Puncture/Hole (1504); Inadequacy of Device Shape and/or Size (1583)
Health Effect - Clinical Codes Pain (1994); Discomfort (2330)
Date of Event 03/06/2026
Type of Reportable Event Serious Injury
Event or Problem Description
It was reported that the patient with this inflatable penile prosthesis (ipp) experienced pain and discomfort inside the corporal bodies.Upon inspection, wear of the outer layer of the cylinders was identified, resulting in cylinder buckling.A potencial oversizing of the device or a weak point in the patient corporal body were the possible causes of the discomfort.As a result, the cylinders and pump were explanted and replaced.No further patient complications were reported.
 
Event or Problem Description
It was reported that the patient with this inflatable penile prosthesis (ipp) experienced pain and discomfort inside the corporal bodies.Upon inspection, wear of the outer layer of the cylinders was identified, resulting in cylinder buckling.A potencial oversizing of the device or a weak point in the patient corporal body were the possible causes of the discomfort.As a result, the cylinders and pump were explanted and replaced.No further patient complications were reported.
 
Additional Manufacturer Narrative
The device is not available for analysis; therefore, no physical analysis of the product could be performed.The reported device performance allegation cannot be confirmed.The device history record (dhr) review confirmed that the device met all material, assembly and performance specifications.A review of the device instructions for use (ifu) was completed and did not reveal any evidence of device misuse, off label use, or failure to follow instructions.A risk review was completed and confirmed that the events of pain/discomfort, excess force or friction on silicone leads to stress, strains, holes or tears, incorrect cylinder size and device cylinder aneurysm/bulge were defined in the product risk documentation.This event type has been accounted for during analysis to support acceptable risk benefit for the product.Based on the information available the cause that contributed to the reported event cannot be established.Therefore, a conclusion code of cause not established was assigned to this investigation.This record was identified during a retrospective review of complaint events; an electronic medical device report (emdr) was not previously submitted per applicable reporting criteria.Boston scientific has opened a nonconforming events and prevention investigation (ncep-00181844) to determine the potential cause for why an emdr was not previously submitted to the food and drug administration (fda) in accordance with boston scientific's established medical device reporting process and fda adverse event reporting requirements.Information gained through these investigation efforts will be utilized to take action(s), as necessary.
 
Additional Manufacturer Narrative
Correction to field g1: mfr site country the device is not available for analysis; therefore, no physical analysis of the product could be performed.The reported device performance allegation cannot be confirmed.The device history record (dhr) review confirmed that the device met all material, assembly and performance specifications.A review of the device instructions for use (ifu) was completed and did not reveal any evidence of device misuse, off label use, or failure to follow instructions.A risk review was completed and confirmed that the events of pain/discomfort, excess force or friction on silicone leads to stress, strains, holes or tears, incorrect cylinder size and device cylinder aneurysm/bulge were defined in the product risk documentation.This event type has been accounted for during analysis to support acceptable risk benefit for the product.Based on the information available the cause that contributed to the reported event cannot be established.Therefore, a conclusion code of cause not established was assigned to this investigation.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
AMS 700 LGX MS Pump
Common Device Name
prosthesis, penis, inflatable
Manufacturer (Section D)
BOSTON SCIENTIFIC CORPORATION
300 boston scientific way
marlborough MA 01752
Manufacturer (Section G)
Boston Scientific Corporation
300 boston scientific way
marlborough MA 01752
Manufacturer Contact
farshad fahimi
4100 hamline avenue north
building c
saint paul, MN 55112
MDR Report Key50714892
Report Number2124215-2026-48421
Device Sequence Number18794516
Product Code JCW
UDI-Device Identifier00878953009810
UDI-Public(01)00878953009810(17)231002(10)1100027977
Combination Product (Y/N)Y
Initial Reporter StateNY
Initial Reporter CountryUS
PMA/510(K) Number
N970012
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2021
Device Explanted Year2026
Reporter Type Manufacturer
Report Source Health Professional,Company Representative
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 09/21/2026
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Lay User/Patient
Device Expiration Date10/02/2023
Device Model Number72404252-10
Device Catalogue Number72404252-10
Device Lot Number1100027977
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 03/06/2026
Supplement Date Received by Manufacturer09/09/2026
Initial Report FDA Received Date09/10/2026
Supplement Report FDA Received Date09/21/2026
Date Device Manufactured10/02/2021
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention; Hospitalization;
Patient Age72 YR
Patient SexMale
-
-