MAUDE Adverse Event Report: Medline Industries, LP MEDLINE; TEST,HCG,PREGNANCY,URINE,CASS, 25 MIU

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

Medline Industries, LP MEDLINE; TEST,HCG,PREGNANCY,URINE,CASS, 25 MIU Back to Search Results
Catalog Number MPH12025HCG
Medical Device Problem Code False Positive Result (1227)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 03/25/2025
Type of Reportable Event Malfunction
Event or Problem Description
It was reported that "2 false positive pregnancy tests out of this box.".
 
Additional Manufacturer Narrative
A complaint was received, and the customer reported receiving medline premium hcg pregnancy tests that "false positive"; the customer reported "2 false positive pregnancy tests out of this box.Exp date 6/30/2026." at this time, a sample(s) was not received for investigation; investigation included review of recent complaint trends/scar activity for the last three trend reports, and retained devices from the reported lot number were tested with hcg-negative clinical urine samples and qc cut-off positive controls.The complaint was not confirmed; no trends were identified for the specified material number and its corresponding defect type, and no false results were observed during in-house testing.A root cause could not be determined from the available information as the reported issue was not replicated during testing of retained product; the division will continue to monitor trends.There were no reports of death, serious injury, medical intervention, or follow up care.It has been determined that the reported event could cause or contribute to serious injury if it were to occur again.In an abundance of caution, this is a reportable event.If additional information becomes available this report will be reopened and reevaluated.This medical device report (mdr) is being submitted as part of retrospective review activities, which include review of historical data in accordance with regulations at 21 cfr part 803.The information included in this mdr reflects the information reasonably known to the manufacturer at the time of this retrospective review and submission.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
MEDLINE
Common Device Name
TEST,HCG,PREGNANCY,URINE,CASS, 25 MIU
Manufacturer (Section D)
Medline Industries, LP
three lakes drive
northfield IL 60093 -275
Manufacturer Contact
sasidhar lakku
three lakes drive
northfield, IL 60093--275
MDR Report Key50714900
Report Number1417592-2026-35754
Device Sequence Number10611635
Product Code JHI
UDI-Device Identifier10888277191877
UDI-Public10888277191877
Combination Product (Y/N)N
Initial Reporter StatePA
Initial Reporter CountryUS
PMA/510(K) Number
K993317
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source Company Representative
Type of Report Initial
Report Date (Section B) 09/07/2026
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Catalogue NumberMPH12025HCG
Device Lot Number0000891076
Was Device Available for Evaluation? Yes
Is the Reporter a Health Professional? No
Was the Report Sent to FDA? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 03/25/2025
Initial Report FDA Received Date09/10/2026
Was Device Evaluated by Manufacturer? (Y/N) No
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Patient SexUnknown
-
-