The account reported finding drape coming apart, and the complaint transcript stated that the extremity drape was coming apart in procedure.No sample or photo was received for the investigation; finished good production records, finished good pack build, and trending were reviewed.The defect was not able to be confirmed as there were no samples or detailed pictures received, and without a sample it is difficult to determine a true root cause for the reported issue.There were no issues with the build that likely contributed to the reported issue, there were 0 additional reported complaints for this issue for the account and component in the past 6 months, and the quality team will monitor this issue for trending purposes.There were no reports of death, serious injury, medical intervention, or follow up care.It has been determined that the reported event could cause or contribute to serious injury if it were to occur again.In an abundance of caution, this is a reportable event.If additional information becomes available this report will be reopened and reevaluated.This medical device report (mdr) is being submitted as part of retrospective review activities, which include review of historical data in accordance with regulations at 21 cfr part 803.The information included in this mdr reflects the information reasonably known to the manufacturer at the time of this retrospective review and submission.
|