Medline Industries, LP MEDLINE; DBD-SAFETY BLOOD COLLECTION NEEDLE 21GX1 Back to Search Results
Catalog Number DYNDSBCN21G
Medical Device Problem Code Material Twisted/Bent (2981)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 04/09/2025
Type of Reportable Event Malfunction
Event or Problem Description
It was reported that "the luer is bending when we try to insert the tube and sometimes when the tube is removed, blood in the needle squirts out.".
 
Additional Manufacturer Narrative
The complainant reported the luer is bending when they try to insert the tube and sometimes when the tube is removed, blood in the needle squirts out.Medline received one case of samples, randomly selected 60 samples from the various lots and boxes of samples, tested using a phlebotomy simulation apparatus with one medline onguard blood collection holder per one needle and 3 vacutainer blood collection tubes, and after each completed collection inspected the luer and needles for any bending as well as making sure each safety mechanism was working properly.Among the 60 samples that were tested, medline observed no bending, no leaking, no failure of safety mechanism, or any other modes of failure; investigation results were not confirmed.Based on the review of the details provided for the complaint investigation, medline was unable to confirm the complaint issue or that the product failed to meet specifications.There were no reports of death, serious injury, medical intervention, or follow up care.It has been determined that the reported event could cause or contribute to serious injury if it were to occur again.In an abundance of caution, this is a reportable event.If additional information becomes available this report will be reopened and reevaluated.This medical device report (mdr) is being submitted as part of retrospective review activities, which include review of historical data in accordance with regulations at 21 cfr part 803.The information included in this mdr reflects the information reasonably known to the manufacturer at the time of this retrospective review and submission.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
MEDLINE
Common Device Name
DBD-SAFETY BLOOD COLLECTION NEEDLE 21GX1
Manufacturer (Section D)
Medline Industries, LP
three lakes drive
northfield IL 60093 -275
Manufacturer Contact
ankitha komirishetty
three lakes drieb
northfield, IL 60093--275
MDR Report Key50715154
Report Number1417592-2026-35772
Device Sequence Number10640532
Product Code FMI
UDI-Device Identifier10193489007879
UDI-Public10193489007879
Combination Product (Y/N)N
Initial Reporter StateAL
Initial Reporter CountryUS
PMA/510(K) Number
K182146
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source Company Representative
Type of Report Initial
Report Date (Section B) 09/05/2026
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Catalogue NumberDYNDSBCN21G
Device Lot Number31823100001
Was Device Available for Evaluation? Yes
Is the Reporter a Health Professional? No
Was the Report Sent to FDA? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 04/09/2025
Initial Report FDA Received Date09/10/2026
Was Device Evaluated by Manufacturer? (Y/N) Yes
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Patient SexUnknown
-
-