| Model Number |
G156 |
| Medical Device Problem Codes |
Signal Artifact/Noise (1036); High impedance (1291); Data Problem (3196)
|
| Health Effect - Clinical Code |
Discomfort (2330)
|
| Date of Event |
08/18/2026
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
It was reported that this cardiac resynchronization therapy defibrillator (crt-d) system was interrogated with a programmer that had an inaccurate date and time group.Ts discussed the programmer would need to be evaluated to resolve this.It was also noted this crt-d system exhibited high out of range pacing impedance measurements on the right atrial (ra) lead.Technical services (ts) noted the atrial tachy response (atr) episode showed non-physiologic noise and discussed it was probable early signs of a lead fracture.It was also noted the patient reported not feeling well recently.The patient was scheduled for an in-clinic interrogation to synch the programmer with the accurate date and time.No additional adverse patient effects were reported.This crt-d remains in service.
|
| |
|
Additional Manufacturer Narrative
|
|
This product investigation is still in progress.This report will be updated when product investigation is complete.
|
| |
|