Lot Number 0039779976
Medical Device Problem Code Contamination (1120)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 08/19/2026
Type of Reportable Event Malfunction
Event or Problem Description
It was reported that in preparation/prior to procedure for a atrial fibrillation (afib) procedure a farawave 2.0 catheter was selected for use, when opened, white substance was found.Procedure completed with another of same device.No patient complications were reported.
 
Event or Problem Description
It was reported that in preparation/prior to procedure for a atrial fibrillation (afib) procedure a farawave 2.0 catheter was selected for use, when opened, white substance was found.Procedure completed with another of same device.No patient complications were reported.
 
Additional Manufacturer Narrative
D2a: common device name: percutaneous cardiac ablation catheter for treatment of atrial fibrillation with irreversible electroporation; reported here as the common device name exceeded the character limit for designated field.
 
Additional Manufacturer Narrative
D2a: common device name: percutaneous cardiac ablation catheter for treatment of atrial fibrillation with irreversible electroporation; reported here as the common device name exceeded the character limit for designated field.Device technical analysis the referenced device was returned for analysis.Visual inspection revealed that there were some white marks and spots in the catheter handle.Device history record review it was confirmed this device met specification prior to distribution and there were no manufacturing deviations which could have contributed to the reported event.The reported event was confirmed.Labeling review based on the information provided, there is no evidence that the device was used in a manner inconsistent with the labelled indications/instructions for use (ifu).Review of the ifu confirmed there was relevant content and sufficient guidance with respect to the circumstances described within this complaint.No updates are required to the ifu as a result of this event.Risk review a risk review performed for the device confirmed that the event of foreign material present in device is a known event defined in the product risk management documentation.This event type has been accounted for during product risk analysis to support acceptable risk benefit for the product.Product risk benefit includes consideration of risk severity and rate, and the overall residual risk of the device is acceptable.Investigation conclusion based on the available information, boston scientific concluded the cause of quality control deficiency was selected based on the identification of white stains on the handle section of the catheter.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
FARAWAVE? NAV
Common Device Name
Percutaneous Cardiac Ablation Catheter For Treatment Of Atrial Fibrillation With Irreversible Electroporation
Manufacturer (Section D)
BOSTON SCIENTIFIC CORPORATION
300 boston scientific way
marlborough MA 01752
Manufacturer (Section G)
Boston Scientific Corporation
302 parkway, global park
la aurora - heredia
CS  
Manufacturer Contact
mary crawford
4100 hamline avenue north
dc a330
saint paul, MN 55112
+1(651)440-0979
MDR Report Key50716779
Report Number2124215-2026-48452
Device Sequence Number9852939
Product Code QZI
UDI-Device Identifier00191506043193
UDI-Public(01)00191506043193(17)270710(10)39779976
Combination Product (Y/N)N
Initial Reporter StateNY
Initial Reporter CountryUS
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source Health Professional,Company Representative
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 10/02/2026
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Lot Number0039779976
Was Device Available for Evaluation? Yes
Date Returned to Manufacturer08/27/2026
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 08/19/2026
Supplement Date Received by Manufacturer09/28/2026
Initial Report FDA Received Date09/10/2026
Supplement Report FDA Received Date10/02/2026
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured07/10/2026
Is the Device Labeled for Single Use? (Y/N) Yes
Usage of Device Initial
Patient Sequence Number1
Patient SexUnknown
-
-