| Lot Number |
0039779976 |
| Medical Device Problem Code |
Contamination (1120)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
08/19/2026
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
It was reported that in preparation/prior to procedure for a atrial fibrillation (afib) procedure a farawave 2.0 catheter was selected for use, when opened, white substance was found.Procedure completed with another of same device.No patient complications were reported.
|
| |
|
Event or Problem Description
|
|
It was reported that in preparation/prior to procedure for a atrial fibrillation (afib) procedure a farawave 2.0 catheter was selected for use, when opened, white substance was found.Procedure completed with another of same device.No patient complications were reported.
|
| |
|
Additional Manufacturer Narrative
|
|
D2a: common device name: percutaneous cardiac ablation catheter for treatment of atrial fibrillation with irreversible electroporation; reported here as the common device name exceeded the character limit for designated field.
|
| |
|
Additional Manufacturer Narrative
|
|
D2a: common device name: percutaneous cardiac ablation catheter for treatment of atrial fibrillation with irreversible electroporation; reported here as the common device name exceeded the character limit for designated field.Device technical analysis
the referenced device was returned for analysis.Visual inspection revealed that there were some white marks and spots in the catheter handle.Device history record review
it was confirmed this device met specification prior to distribution and there were no manufacturing deviations which could have contributed to the reported event.The reported event was confirmed.Labeling review
based on the information provided, there is no evidence that the device was used in a manner inconsistent with the labelled indications/instructions for use (ifu).Review of the ifu confirmed there was relevant content and sufficient guidance with respect to the circumstances described within this complaint.No updates are required to the ifu as a result of this event.Risk review
a risk review performed for the device confirmed that the event of foreign material present in device is a known event defined in the product risk management documentation.This event type has been accounted for during product risk analysis to support acceptable risk benefit for the product.Product risk benefit includes consideration of risk severity and rate, and the overall residual risk of the device is acceptable.Investigation conclusion
based on the available information, boston scientific concluded the cause of quality control deficiency was selected based on the identification of white stains on the handle section of the catheter.
|
| |
|