• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: BOSTON SCIENTIFIC CORPORATION FARAWAVE? NAV; Percutaneous Cardiac Ablation Catheter For Treatment Of Atrial Fibrillation With Irreversible Electroporation

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BOSTON SCIENTIFIC CORPORATION FARAWAVE? NAV; Percutaneous Cardiac Ablation Catheter For Treatment Of Atrial Fibrillation With Irreversible Electroporation Back to Search Results
Lot Number 0038835599
Medical Device Problem Code Air/Gas in Device (4062)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 08/19/2026
Type of Reportable Event Malfunction
Event or Problem Description
It was reported that a pulse field ablation (pfa) procedure was being performed using a farawave catheter and a faradrive steerable sheath clear.After the farawave catheter was inserted into the faradrive sheath, air aspiration was performed within the sheath.Despite repeated attempts, air continued to be aspirated from the catheter tip.Following sheath removal, the patient experienced st-segment elevation, hypotension, and atrioventricular (av) block.Noradrenaline was administered, and coronary angiography (cag) was performed.Vasospasm was observed during cag, and nitroglycerin was administered.Although vasospasm occurred, the relationship between the event and the device could not be determined.After the patient's condition stabilized, cag reportedly demonstrated air ingress.The procedure was resumed, and both the farawave catheter and faradrive steerable sheath clear were replaced with a second farawave catheter and a second faradrive steerable sheath clear.Air aspiration was performed again; however, the same issue of continued air aspiration was reproduced.At the physician's discretion, the farawave catheter was removed and replaced with a non-boston scientific pfa catheter in order to continue the procedure.Pulmonary vein isolation (pvi) was subsequently completed, and the procedure was finished successfully.The physician reported that because the issue persisted despite replacement of both the catheter and sheath, the event was not believed to be product related.The patient remained stable following the procedure and was discharged from the hospital.
 
Additional Manufacturer Narrative
D2a: common device name: percutaneous cardiac ablation catheter for treatment of atrial fibrillation with irreversible electroporation; reported here as the common device name exceeded the character limit for designated field.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
FARAWAVE? NAV
Common Device Name
Percutaneous Cardiac Ablation Catheter For Treatment Of Atrial Fibrillation With Irreversible Electroporation
Manufacturer (Section D)
BOSTON SCIENTIFIC CORPORATION
300 boston scientific way
marlborough MA 01752
Manufacturer (Section G)
Boston Scientific Corporation
302 parkway, global park
la aurora - heredia
CS  
Manufacturer Contact
mary crawford
4100 hamline avenue north
dc a330
saint paul, MN 55112
+1(651)440-0979
MDR Report Key50716940
Report Number2124215-2026-48459
Device Sequence Number18800725
Product Code QZI
UDI-Device Identifier00191506043193
UDI-Public(01)00191506043193(17)270314(10)38835599
Combination Product (Y/N)N
Initial Reporter CountryJA
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source Other,Foreign,Health Professional,Company Representative
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 09/30/2026
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Lot Number0038835599
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 08/21/2026
Supplement Date Received by Manufacturer09/28/2026
Initial Report FDA Received Date09/10/2026
Supplement Report FDA Received Date09/30/2026
Date Device Manufactured03/14/2026
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
FARADRIVE Sheath (BSC)
Patient SexUnknown
-
-