| Model Number |
A42011A |
| Medical Device Problem Code |
Use of Device Problem (1670)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
06/18/2026
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
The customer returned the reusable endoscope sheath for a separate issue.During the device analysis, it was observed that the ceramic tip was broken was found.There was no patient involvement.
|
| |
|
Additional Manufacturer Narrative
|
|
The initial evaluation findings were confirmed.Based on the results of the investigation, it is likely the following led to the malfunction: it was found that the ceramic tip (insulation beak) was broken and repaired by a third-party during the service inspection.Olympus will continue to monitor the field performance of this device.Should any additional relevant information become available, a supplemental report will be submitted.
|
| |
|
Search Alerts/Recalls
|