• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: MIS Implants Technologies Ltd. C1 coni. con. implant D3.75 L8mm, SP; implant, endosseous, root-form

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

MIS Implants Technologies Ltd. C1 coni. con. implant D3.75 L8mm, SP; implant, endosseous, root-form Back to Search Results
Lot Number W23013237
Medical Device Problem Code Loss of Osseointegration (2408)
Health Effect - Clinical Code Insufficient Information (4580)
Type of Reportable Event Serious Injury
Event or Problem Description
It was reported that a patient experienced a dental implant loss.
 
Additional Manufacturer Narrative
Since this event resulted in a serious injury, it is reportable per 21 cfr part 803.The device was not evaluated because the issue is a known inherent risk of the device.We will continue to track and monitor the trend.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
C1 coni. con. implant D3.75 L8mm, SP
Common Device Name
implant, endosseous, root-form
Manufacturer (Section D)
MIS Implants Technologies Ltd.
p.o. box 7
bar lev industrial park 20156 00
IS  2015600
Manufacturer Contact
mr. dan eagar
suite 60w
york, PA 17401
MDR Report Key50717579
Report Number3004203816-2026-25507
Device Sequence Number10586113
Product Code DZE
UDI-Device Identifier07290108992076
UDI-Public07290108992076
Combination Product (Y/N)N
Initial Reporter CountryRS
PMA/510(K) Number
K112162
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Initial Reporter Occupation Non-Healthcare Professional
Type of Report Initial
Report Date (Section B) 09/10/2026
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Device Lot NumberW23013237
Was Device Available for Evaluation? Yes
Is the Reporter a Health Professional? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 08/11/2026
Initial Report FDA Received Date09/10/2026
Was Device Evaluated by Manufacturer? (Y/N) No
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
Yes
Usage of Device Unknown
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
Patient Age51 YR
Patient SexUnknown
-
-