| Model Number |
GF-UCT180 |
| Medical Device Problem Codes |
Detachment of Device or Device Component (2907); Scratched Material (3020); Material Split, Cut or Torn (4008)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
08/04/2026
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
The customer reported that the ultrasound gastrovideoscope had a separate issue.During the initial evaluation of the device, missing thixotropic adhesive (ke-45) on the distal end (c-body) forceps cover, a cut pink rubber and scratches on the light guide lens cement were observed.There was no patient involvement.
|
| |
|
Additional Manufacturer Narrative
|
|
Based on the final investigation of the device the initial evaluation findings were confirmed.A root cause could not be identified.Based on the results of the investigation, the most probable cause of the malfunction is traced to component failure.Olympus will continue to monitor the field performance of this device.Should any additional relevant information become available, a supplemental report will be submitted.
|
| |
|