The customer reported that the needles are bent and the safety doesn¿t cover the needle; no medical intervention, serious injury, or follow-up care has been reported.Medline received samples in new condition, tested 25 samples total, and performed visual inspection and functional testing using an injection pad.The visual inspection revealed no manufacturing defects preventing the samples from working as designed, all 25 needles passed the functional test, and the complaint was not confirmed.The cause was documented as unconfirmed defect - not confirmed with received sample, and the division will continue to monitor the issue for trending.There were no reports of death, serious injury, medical intervention, or follow up care.It has been determined that the reported event could cause or contribute to serious injury if it were to occur again.In an abundance of caution, this is a reportable event.If additional information becomes available this report will be reopened and reevaluated.This medical device report (mdr) is being submitted as part of retrospective review activities, which include review of historical data in accordance with regulations at 21 cfr part 803.The information included in this mdr reflects the information reasonably known to the manufacturer at the time of this retrospective review and submission.
|