|
Product complaint # (b)(4)
this report is being submitted pursuant to the provisions of 21 cfr, part 803 (and/or part 4, as applicable).This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by depuy synthes, or its employees that the report constitutes an admission that the product, depuy synthes, or its employees caused or contributed to the potential event described in this report.Additional narrative:
h3, h6: product investigation summary:
the product was returned to depuy synthes for evaluation.Visual analysis of the summit universal broach handle found signs consistent with heavy and constant usage in a clinical setting over time; such as multiple impaction marks, deformation and washing stains.A functional test performed with a mating device revealed looseness in the lever of the locking mechanism, even before the assembling process.It is suspected that an internal locking mechanism component was worn due to repeated use.The observed condition was identified as an end of life indicator; damage consistent with repeated use and servicing.The lifecycle requirements of the device are event related and depend on the use and inspection of the device in clinical practice.As the device can be damaged on the first or 100th use, the device must be properly inspected prior to each surgical use.Refer to the device/country specific ifu for information related to end of life, reprocessing instructions, and inspection procedures.A dimensional inspection was not performed since it was not applicable to the complaint condition.The overall complaint was confirmed as the observed condition of the summit universal broach handle would have contributed to the complained device issue.¿
based on the investigation findings, it has been determined that no corrective and/or preventative action is proposed.There is no indication that a design or manufacturing issue has caused the reported complaint condition.As part of depuy synthes quality process, all devices are manufactured, inspected, and released to approved specifications.Additional monitoring for any potential safety signals will be conducted through complaint trending and other post-market safety surveillance activities.Device history lot: a device history record (dhr) review or manufacturing records evaluation (mre) was not possible because the date code (ak0605) provided is not a valid finished goods lot#
if information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
|