| Model Number |
STQ4-RCV-A0 |
| Medical Device Problem Code |
Adverse Event Without Identified Device or Use Problem (2993)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
07/28/2026
|
|
Type of Reportable Event
|
Death
|
|
Additional Manufacturer Narrative
|
|
The other adverse events issues questionnaire was reviewed.However, there were no alleged deficiencies as therapy was working properly and the death was unrelated to the curonix device.The transmitter assemblies associated with the complaint were returned.Although the transmitter assemblies were working properly and there were no alleged deficiencies, the investigation found a damaged component as l12 was broken on both devices.The damages observed are considered cosmetic and did not effect the functionality of the device.The stimulator is used to treat pain.The cause of the patients death is unrelated to the curonix device.The provided information does not evidence that the curonix device contributed to the issue (no fault found).Rates reviewed at the most recent complaint trending meeting do not indicate a significant increase in other adverse events issues.Other adverse events issues rates remain acceptably low; thus, a capa is not required.Other adverse events issues rates will continue to be tracked and trended.Refer to capa-2024-0020 for additional investigation details for the damaged component l12.
|
| |
|
Event or Problem Description
|
|
Curonix support was notified the patient had passed on (b)(6) 2026 from non-device related factors.There were no alleged deficiencies, therapy was working properly, and the family would be returning the transmitter assemblies.
|
| |
|