It was reported that during annual service, it was observed that the fms vue pump device had a mechanical problem, it could not swing back.The "blood stop", "fill chamber" and "shaver suction setting" leds could not be illuminated.This event did not occur during surgery. there were no reports of injuries, medical intervention or prolonged hospitalization. the exact date of the event was unknown.However, it was reported that the event occurred in 2026.All available information has been disclosed. .
|
|
This report is being submitted pursuant to the provisions of 21 cfr, part 803.This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by depuy synthes, or its employees that the report constitutes an admission that the product, depuy synthes, or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial report, a follow-up report will be filed as appropriate.H11 additional narrative:
the complaint device was received at the service center then, the device was forwarded to oem, and the following defects were identified:
functional : screen issue/display issue
during the service, identified the follower arm on suction pump head has mechanical problem, can't swing back."blood stop", "fill chamber" and "shaver suction setting" leds can¿t be illuminated.Parts are obsolete, unavailable to order.Discarded the unrepaired unit as per customer request.Per service reports, this complaint was confirmed.Based on the investigation findings, the potential root cause is traced to faulty parts.It has been determined that no corrective and/or preventative action is proposed and there is no indication that a design or manufacturing issue has caused the reported complaint condition.As part of depuy synthes quality process, all devices are manufactured, inspected, and released to approved specifications.Additional monitoring for any potential safety signals will be conducted through complaint trending and other post-market safety surveillance activities.A manufacturing record evaluation was performed for the finished device, and there was a non-conformance unrelated to the reported complaint.
|