• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: Institut Straumann AG BLC, ? 4.0mm RB, SLActive 12mm, RXD; Endosseous Dental Implant

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

Institut Straumann AG BLC, ? 4.0mm RB, SLActive 12mm, RXD; Endosseous Dental Implant Back to Search Results
Catalog Number 035.9312S
Medical Device Problem Code Separation Failure (2547)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 08/04/2026
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
The batch number could be verified.Our manufacturing q-system assures,\ that production and process controls are in place to ensure that batches\ confirm to the applicable specifications before they are distributed.?\ the removal of a dental implant during surgery without the replacement\ of another dental implant is a known inherent risk of the procedure due\ to either lack of primary stability of the implant (patienor procedure\ related).It may also include the removal of an implant after\ osseointegration due to either the clinician's or patient¿s decision.\ the manufacturer¿s trend analysis confirms that usually procedural\ errors and/or patient's condition contribute to the event.
 
Event or Problem Description
The clinician reports that the day the implant was placed in ada 11, failure occurred upon insertion.Details of surgery: primary stability not achieved and implant surface not completely covered with bone.Patient presented with bone type iii.The device was forwarded to the manufacturer.There were no reported patient injuries or complications.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
BLC, ? 4.0mm RB, SLActive 12mm, RXD
Common Device Name
Endosseous Dental Implant
Manufacturer (Section D)
Institut Straumann AG
peter merian-weg 12
basel 04002
SZ  04002
Manufacturer (Section G)
Institut Straumann AG
peter merian-weg 12
basel 04002
SZ   04002
Manufacturer Contact
jennifer jackson
60 minuteman road
andover, MA 01810
+1(978)747-2509
MDR Report Key50719477
Report Number0009613348-2026-014073
Device Sequence Number10066007
Product Code DZE
UDI-Device Identifier07630031780096
UDI-Public07630031780096
Combination Product (Y/N)N
Initial Reporter StateNY
Initial Reporter CountryUS
PMA/510(K) Number
K230108
Summary Report (Y/N)N
Device Implanted Year2026
Device Explanted Year2026
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Foreign
Initial Reporter Occupation Dentist
Type of Report Initial
Report Date (Section B) 09/10/2026
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Catalogue Number035.9312S
Device Lot NumberYEAR5
Was Device Available for Evaluation? Yes
Date Returned to Manufacturer08/17/2026
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Event Location Other
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 09/10/2026
Initial Report FDA Received Date09/10/2026
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured04/29/2026
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
Patient Age58 YR
Patient SexMale
-
-