The customer reported a quality issue: 1 pack was visibly wet throught the wrapper and sterility was compromised.A sample pack was received and reviewed; the back table cover that wrapped the pack was observed to be wet, the pack was opened and the components and subassemblies within were also wet, and component 52011 - saline str 100ml f/irrig had no longer any more liquid within.The reported issue was confirmed.Based on the complaint details the root cause is considered to be a vendor product issue relating to the component manufacturing process of component 52011 - saline str 100ml f/irrig (seal) that had leaked within the pack; the vendor determined the defect to be from a manufacturing error, and the division will continue to monitor this issue for trending.There were no reports of death, serious injury, medical intervention, or follow up care.It has been determined that the reported event could cause or contribute to serious injury if it were to occur again.In an abundance of caution, this is a reportable event.If additional information becomes available this report will be reopened and reevaluated.This medical device report (mdr) is being submitted as part of retrospective review activities, which include review of historical data in accordance with regulations at 21 cfr part 803.The information included in this mdr reflects the information reasonably known to the manufacturer at the time of this retrospective review and submission.
|