| Model Number |
Cardiac Workstation 7000 |
| Medical Device Problem Code |
Incorrect Measurement (1383)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
08/19/2026
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
Philips received a complaint on a cardiac workstation 7000 indicating that there was an issue with pins showing inaccuracy in functionality.The device was not in clinical use at the time of the event.No adverse event was reported.
|
| |
|
Additional Manufacturer Narrative
|
|
Analysis was performed remote service personnel which included troubleshooting.Based on the results of the analysis, it was determined that the product has malfunctioned and the cause of the reported problem was a faulty patient interface module (pim).The issue was resolved by replacing the pim and the product is back in service.
|
| |
|