Model Number AP14R324S
Medical Device Problem Code Material Separation (1562)
Health Effect - Clinical Code Foreign Body In Patient (2687)
Date of Event 08/17/2026
Type of Reportable Event Serious Injury
Event or Problem Description
It was reported that an asahi gladius mongo 14 es guide wire was used during a plain old balloon angioplasty (poba) to treat a heavily calcified, heavily tortuous chronic total occlusion (cto) in the mid anterior tibial artery (ata).The ata was totally occluded throughout its entire length from the proximal segment.The gladius mongo 14 es guide wire was advanced into the proximal ata.While advancing the tip of the guide wire to make a small j shape, it got caught and entangled, forming a knot in one of the side branches of the ata branching off the proximal segment.Subsequently, the entangled portion of the guide wire could not be unknotted and withdrawn.The tip of the gladius mongo 14 es guide wire broke off within the small side branch of the ata.After the tip fracture, the proximal segment of the gladius mongo 14 es guide wire was successfully withdrawn from the patient anatomy.The separated wire tip remained in the patient anatomy and could not be retrieved.The physician commented that the tip of the guide wire became entangled and formed a knot as a result ofrapid advancement and rotation of the wire within an extremely narrow side branch of the ata.The size of the knot was presumed to be less than 3 mm.It was informed that there were no adverse patient effects associated with this event and the patient was discharged after the procedure.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ASAHI Gladius Mongo 14 ES
Common Device Name
Peripheral Vascular Guide Wire
Manufacturer (Section D)
Asahi Intecc CO Ltd
3-100 akatsuki-cho
seto, aichi 489-0 071
JA  489-0071
MDR Report Key50722714
Report Number3004718255-2026-00038
Device Sequence Number18719367
Product Code DQX
UDI-Device Identifier04547327123747
UDI-Public(01)04547327123747(17)290131(10)260202A111
Combination Product (Y/N)N
Initial Reporter StateFL
Initial Reporter CountryUS
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Voluntary
Initial Reporter Occupation Physician
Type of Report Initial
Report Date (Section B) 09/10/2026
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model NumberAP14R324S
Device Lot Number260202A111
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? Yes
Event Location Hospital
Initial Date Received by Manufacturer Not provided
Initial Report FDA Received Date09/10/2026
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Outcome Attributed to Adverse Event Disability;
Patient Age60 YR
Patient SexMale
Date Report Sent to FDA09/10/2026
-
-