It was reported that device embolization occurred.The patient required surgical removal.A left atrial appendage (laa) closure procedure was being performed under transesophageal echocardiography (tee) guidance.Pre-procedure imaging demonstrated challenging laa anatomy.Transseptal puncture was performed using a fixed versacross transseptal system.A 27mm watchman flx pro delivery system and closure device (wds) was initially attempted but was displaced proximally due to the appendage anatomy.A 24mm wds was then attempted but did not achieve adequate compression.The physician subsequently returned to a 27mm wds using a watchman trusteer access system (was).After several positioning attempts, the 27mm wds met release criteria and was released.Following release, the closure device appeared to have shifted.The physician elected to retrieve the closure device and abort the procedure.After recrossing the septum and advancing a biotome for retrieval, the sheath contacted the closure device, causing it to dislodge and embolize into the left ventricular outflow tract (lvot).Percutaneous retrieval from the lvot was attempted but was unsuccessful.Cardiovascular surgery was consulted, and the patient was taken to the operating room, where the closure device was surgically removed and the laa was clipped.The patient was transferred to recovery and is expected to fully recover.It was further reported the patient was kept in the hospital for several days following the surgery then discharged to a rehab facility to help with recovery.
|