| Model Number |
72400024 |
| Medical Device Problem Codes |
Fluid/Blood Leak (1250); Material Puncture/Hole (1504)
|
| Health Effect - Clinical Code |
Urinary Incontinence (4572)
|
| Date of Event |
08/18/2026
|
|
Type of Reportable Event
|
Serious Injury
|
|
Event or Problem Description
|
|
It was reported that the patient with an artificial urinary sphincter (aus) experienced recurring incontinence.During revision surgery, the pressure regulating balloon (prb) was identified to have a fluid loss, and a pinhole was observed near the prb hub during back-table testing.A revision procedure was performed during which the prb was explanted and replaced with a new one.No patient complications were reported.
|
| |
|
Additional Manufacturer Narrative
|
|
Model number/catalog number: 72404127.Serial number: n/a.Batch/lot number: 1100498789.Model/catalog description: control pump fgs iz.Unique identifier (udi) #: (b)(4).Model number/catalog number: 72404138.Serial number: n/a.Batch/lot number: 1100255817.Model/catalog description: belt cuff fgs 8.0 cm iz.Unique identifier (udi) #: (b)(4).
|
| |
|