• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: Abbott Medical (CRM-Sylmar) ENDURITY SR; Pulse Generator, Permanent, Implantable

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

Abbott Medical (CRM-Sylmar) ENDURITY SR; Pulse Generator, Permanent, Implantable Back to Search Results
Model Number PM1160
Medical Device Problem Code Adverse Event Without Identified Device or Use Problem (2993)
Health Effect - Clinical Codes Wound Dehiscence (1154); Unspecified Infection (1930); Pocket Erosion (2013)
Date of Event 08/14/2026
Type of Reportable Event Serious Injury
Event or Problem Description
It was reported that the patient presented for a post-operative follow-up.It was noted that the patient repeatedly developed pocket infections associated with redness, swelling, and ulceration.The wound remained open, with the pacemaker exposed, despite two debridement procedures.A small portion of heart tissue was observed on the lead.The physician believed that the infection was device and procedure related.The entire system, including the pacemaker, and the right ventricular lead, was explanted and replaced with a leadless pacemaker.The patient remained in stable condition.
 
Additional Manufacturer Narrative
A device history record (dhr) review was performed and review of the sterilization records confirmed normal sterilization cycles for the products.The device met specifications prior to leaving abbott manufacturing facilities.The results of the investigation are inconclusive since the device was not returned for analysis.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ENDURITY SR
Common Device Name
Pulse Generator, Permanent, Implantable
Manufacturer (Section D)
Abbott Medical (CRM-Sylmar)
15900 valley view court
sylmar CA 91342
Manufacturer (Section G)
Abbott Medical (CRM-Sylmar)
15900 valley view court
sylmar CA 91342
Manufacturer Contact
navpreet singh
15900 valley view court
sylmar, CA 91342
MDR Report Key50725406
Report Number2017865-2026-19254
Device Sequence Number10587140
Product Code NVZ
UDI-Device Identifier05414734507042
UDI-Public(01)05414734507042(10)P000253989(17)270930
Combination Product (Y/N)N
Initial Reporter CountryCH
PMA/510(K) Number
P880086
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2026
Device Explanted Year2026
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Foreign,Health Professional,User Facility,Company Representative
Initial Reporter Occupation Physician
Type of Report Initial
Report Date (Section B) 09/11/2026
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Model NumberPM1160
Device Lot NumberP000253989
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 08/23/2026
Initial Report FDA Received Date09/11/2026
Was Device Evaluated by Manufacturer? (Y/N) No
Date Device Manufactured04/11/2026
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
Patient SexUnknown
-
-